XR-100SDD Silicon Drift Detector SDD-Obsolete
The SDD is now Obsolete, please see our FAST SDD.
The XR-100SDD is a thermoelectrically cooled solid-state silicon drift detector (SDD) and preamplifier. It is recommended for applications requiring the best energy resolution, very high-count rates, and lowest X-ray energies. Its performance, small size, and low cost make it the ideal detector for many laboratory and OEM X-ray spectroscopy applications, including EDS and XRF.

Product Details
The XR-100SDD can provide a resolution of 125 eV FWHM at the Mn Kα line (electronic noise of 4.5 electrons rms), a peak to background of 20,000:1, an output count rate over 500 kcps, and can detect X-rays down to the Be Kα line (110 eV). It has a 25 mm2 active area and is 500 μm thick.
The standard XR-100SDD has a 0.5 mil Be window for good efficiency above 2 keV. For lower energies, the optional Patented C-Series windows available with our FAST SDD® provide good efficiency down to the B line. For count rates >200 kcps, the FAST SDD® is recommended.
In the XR-100SDD, the detector is mounted on an extender (several different lengths are available), with the preamplifier in the attached metal box. It requires a separate signal processor and power supplies; Amptek’s PX5 is recommended and is ideally suited for most laboratory uses. The same SDD detectors are available in the smaller X-123SDD package or with smaller preamplifiers for OEMs and custom systems.
Amptek recently brought silicon wafer manufacturing in-house and improved the process. The result is a detector with lower noise, lower leakage current, better charge collection, and uniformity from detector to detector.
- Lower noise → Better resolution down to 125 eV FWHM
- Lower leakage current → Higher temperature operation (save battery life)
- Better charge collection → Better photopeak shape (no tailing)
- Quality → Detectors have consistent performance allowing for easier calibrations
The XR-100SDD is an enhanced version of Amptek’s thermoelectrically cooled X-ray detectors. It uses a silicon drift detector (SDD), a silicon photodiode with a special electrode configuration giving very low capacitance, resulting in low electronic noise at high frequencies. This provides the improved energy resolution and count rate. The SDD uses special “drift” electrodes to guide the charge into its anode, hence the name “drift detector”.
As with Amptek’s other XR-100 detectors, the photodiode is mounted on a two stage thermoelectric cooler, keeping the detector and its input JFET at approximately -55 °C, reducing electronic noise without cryogenic liquid nitrogen. This cooling permits high performance in a compact, convenient package, and has been a core enabler of the portable XRF analyzer.
The hermetic TO-8 package of the detector has a light tight, vacuum tight, thin Be window to enable soft X-ray detection. There is vacuum inside the enclosure for optimum cooling. The XR-100SDD detector includes an internal multilayer collimator to minimize background and spectral artifacts. It has a reset-style preamplifier.
In the XR-100SDD, the preamplifier is enclosed in a metal box, 3.0 x 1.75 x 1.125 in (76.2 x 44.45 x 28.58 mm), with the detector on an extender (available lengths range from 3/8” to 9” or 9.53 to 228.6 mm). The XR-100SDD with a 5” or 9” extender is suitable for vacuum measurements, using the optional CP75 vacuum flange. Alternate preamplifiers are available, recommended for OEMs or where space is limited.