MCA-8000D Option PA for Airborne and Suspended Particle Applications

Digital Multichannel Analyzer

The Option PA package has been developed to facilitate the use of the MCA-8000D for particle counting in airborne (Size Calibration) and liquid suspended (Number Calibration) particle applications. 

MCA8000D Pocket MCA digital multichannel analyzer

Product Details

The Option PA package has been developed to facilitate the use of the MCA-8000D for particle counting in airborne (Size Calibration) and liquid suspended1 (Number Calibration) particle applications. The unit is calibrated and certified traceable to the National Institute of Standards and Technology (NIST). The Option PA package is capable of detecting pulses from 5 mV to 10 V. The MCA-8000D is typically connected to the output of a particle sensor. It detects and displays a spectrum of pulse heights allowing the user to determine if a given particle size is producing the correct voltage. The software included with the MCA provides information on the peak center (centroid and mean calculation) making it easy to determine if the peak is in the correct position. The MCA is internally calibrated to convert the MCA channel scale to a mV scale and therefore loads automatically, making it unnecessary to load calibration files manually.

Features & Specifications

  • Pulse detection from 5 mV to 10 V
  • 2 Voltage scales: 0-1 V and 0-10 V
  • 100 k input impedance
  • NIST Traceable calibration
  • Certificate of calibration
  • Absolute or first peak detection mode (see figure 1)
MCA peak detection and pulse height analysis diagram
Figure 1.

Tolerances

V ScaleOffsetGain
0-1 V Scale±0.2 mV ±0.2%
0-10 V Scale±2.0 mV±0.5%

The Option PA Package is NOT recommended for standard nuclear instrumentation applications.

References

1Sommer, H.T. “IMPLEMENTING PARTICLE COUNTER CALIBRATION PER ISO 11171-1999.”
TEAM Service, Inc., P.O. Box 220, Merlin, OR 97532, (541)476-4744, HolgerTSo@aol.com
Copyright Society of Automotive Engineers

Downloads

Documentation

H.T. Sommer - Implementing Particle Counter Calibration

PDF

Manual

MCA-8000D Option PA Manual

PDF
Amptek X-ray product technical support

Need Further Information?

Our team is available to assist. Submit your enquiry using the link below, and we will provide the information you require.