Patented Low Energy C-Series Windows

Amptek Patented “C-Series” X-ray windows utilize silicon nitride (Si3N4) with an aluminum coating to extend the low energy response of our silicon drift detectors (SDDs) down to boron (B); they are available exclusively with our FastSDD®.

C-Series detector window options for X-ray spectroscopy

Product Details

Amptek provided an alternative to liquid nitrogen when it introduced the first thermoelectrically cooled detector. Now with the Patented “C-Series,” we offer an alternative to the Beryllium (Be) window for general XRF analysis, and superior performance as compared to polymer windows for soft X-ray analysis.

C2 window transmission test setup in water sample

Applications

C1 Windows: Laboratory, benchtop and handheld instruments. This window is light-tight so it can be used in normal ambient room light. C2 Windows: Vacuum applications. The C2 window has much better efficiency at the lowest energies but is not light-tight and therefore must be used in dark environments.

LEW detector window transmission efficiency versus energy
FAST SDD detector focal plane array assembly

Performance Gallery

FAQs

The Patented C-Series windows offer less attenuation (higher intrinsic efficiency) than the standard Be windows for X-ray energies below 1.5 keV (for elements below Al). The C windows are about a factor of 2 more efficient for Na Kα X-rays (at 1.04 keV) than the thinnest Be window (8 μm). With Be windows it is virtually impossible to measure elements below Na in the periodic table, and even Na and Mg are marginal. The Patented C-Series windows make it practical to measure light elements. 

 

One should measure in vacuum to take full advantage of the Patented C-series windows. Attenuation in air can be more important than attenuation in the windows for low energies. For Na Kα X-rays at 1.04 keV the attenuation length in air is only 2.2 mm: an air path of 2.2 mm will attenuate these X-rays as much as a Be window. This not only reduces the sensitivity at Na, but path length variations become very important: an additional 0.2 mm of air changes the Na Kα intensity by 10%, making accurate analysis difficult.

Specifications

C Series Windows Specification

PDF

Troubleshooting

C Series Window Protection Advice

PDF