Home PageProductsPrice ListLinks & PDFs 
xr100cr x-ray detector

PX2CR: Power Supply & Shaping Amplifier for the XR-100CR

xr-100cr x-ray detector

General

Input AC power to the PX2CR is provided through a standard IEC 320 plug (110/250 VAC, 50-60 Hz). See Figure 1.

The four (4) DC Voltages needed to operate the XR100CR are supplied through a female 9-Pin D-Connector on the PX2CR. The Pin list to this connector is given below. The multiconductor cable which connects the PX2CR to the XR-100CR is provided with the system.

Physical

Size6 x 6 x 3.5 in (15.3 x 15.3 x 8.9 cm)
Weight2.5 lbs (1.15 kg)

9-Pin D-Connector

Pin 1+9 V Preamp power
Pin 2-9 V Preamp Power
Pin 31-Stage Cooler Power 0 to +2 V @ 700 mA maximum
2-Stage Cooler Power 0 to +3 V @ 350 mA maximum
Pin 4+9 V Temperature Monitor Power
Pin 5+H.V. Detector Bias, +100 - 130 V maximum
Pin 6Ground and case
Pin 7Cooler power return
Pin 8Ground and case
Pin 9Ground and case

PX2CR Shaping Amplifier Specifications

PolarityPositive unipolar
Shaping TimeFixed 6 µs, 12 µs, or 20 µs
Pulse WidthSee Figure 2.
Shaping Type7 pole "Triangular" with base line restoration, rise time discrimination (RTD), and pileup rejection.
Sensitivity0 to 1 V/keV (10 turn pot)
Gain0 to X1000
Gain ShiftSee figure 4
Output Impedance< 1 ohm

The output pulse produced by the PX2CR Shaping Amplifier is optimum for most applications using the Si-PIN photodiode detectors, and can be connected directly to the input of a Multichannel Analyzer (MCA). For optimum portability and versatility, use the Amptek MCA8000A "Pocket MCA" with 16k data channels.

Signal Connections

Input from XR100CRFront Panel BNC
Output to MCAFront Panel BNC
Pileup Rejection (PU)Rear Panel BNC, Positive TTL
For the duration of this output gate, any detected pulse must be rejected by the MCA
Input Count Rate (ICR)Rear Panel BNC, Positive TTL <2 µs
When connected to a counter the ICR countrate corresponds to the total number of X-Ray events that strike the detector.

This diagram shows the internal connections between the AXRCR hybrid sensor and the electronics within the case, as well as the external connections to the PX2CR.

XR-100CR x-ray detector system block diagram
Figure 1. XR-100CR Connection Diagram

PX2CR amp out pulse shape
Figure 2. PX2CR Amplifier Output (12µs shaping time)
Shaping Time ConstantPulse Width
Standard 12 µs22 µs FWHM
Optional 6 µs15 µs FWHM
Optional 20 µs54 µs FWHM
XR-100CR x-ray detector output vs. input count rate
Figure 3. Output vs. Input Rate for different Shaping Time Constants
XR-100CR x-ray detector peak shift/resolution vs. count rate
Figure 4. Resolution and Peak Shift vs. Count Rate