xrf-fp

Introduction

The purpose of XRF analysis with Fundamental Parameters (FP) is to convert elemental peak intensities to elemental concentrations and/or film thicknesses. This is achieved typically though a calibration step, where the XRF response function (related to parameters that are independent of the sample matrix) for each element is measured using a known standard of some kind. In some circumstances the analysis may be purely based upon theoretical equations, and the fundamental-parameter database, without any need for a calibration step. The latter is possible for analyses of simple bulk or single-layer thin-film samples, when the film thickness is known (e.g., a bulk sample).

xrf-fp main display
Figure 1. XRF-FP Main display window. Shows the element table with the various parameters and chemical concentrations.

As outlined above, there are only two steps in XRF analysis, whether or not the FP method is used. The first step is to calibrate the response function for each element from one or many standards (called the “Calibration” step). The second step is to produce the sample analysis of a given material, using the previously stored calibration coefficients, and the FP algorithms given a definition of the sample (i.e., number of layers, and which elements are in which layers).

The software will support single layer or bulk composition and thickness analysis of up to 30 elements, calculated as either elements and/or compounds. Up to 4 or more excitation “conditions” are allowed per analysis. Each condition describes a separate analysis, and can be freely defined with any combination of experimental conditions, such as kV, tube anode, filter, detector filter, environment (air, vacuum, He) and acquire time. This allows the analyst to measure some elements with one condition, and others completely differently, such that each analysis can be optimized for the specific element, or group of elements. Likewise, the spectrum processing steps can also be freely defined, and are all part of the so-called condition code setup.

The FP analysis software will support a single or multiple standard calibration scheme, or completely standardless analysis if the tube, detector, environmental and geometry parameters are known. Calibration standards should be passed one at a time and the merging of the calibration standard information is handled internally. After each calibration step, a set of calibration coefficients and associated information, for each of the defined elements, is returned, which can immediately be used if only one standard is employed. When using multiple calibration standards, all the coefficients are merged into one set, and then this final set is available for subsequent quantitative analysis.

The layer thicknesses must be fixed for standardless analysis. Results can be normalized to any value, and MUST be normalized for standardless analysis or when the layer thicknesses are calculated. Elements (or compounds) can be calculated, fixed, or determined by difference. Elements can also be determined by stoichiometry from the compound formulae. Composition results can be calculated in units of Wt% or ppm, and for thin films, units such as ug/cm2 and mg/cm2 are used for mass thickness. The latter can be converted to thickness (microns, microinches, nm, etc.) if the density is known. The density may be input or optionally calculated theoretically.

All the appropriate FP calculations are made both during calibration and for Quantitation, using calculations based upon the Sherman equation. Tube spectra, required for the direct fluorescence calculations, can be supplied by the user, or calculated from built-in models (Ebel, Pella et al.). These tube spectra can be convolved with experimental transfer functions to derive the expected tube spectrum passing through an optic such as a polycapillary bundle. The presence of air paths will also be calculated from the input geometry parameters, for both the source and detector paths. Single-element filters can also be inserted between the tube and the sample, or between the sample and the detector, and the software can accommodate both.

The detector parameters (window, thickness, etc.) will also be used to calculate the various absorption and efficiency effects when x rays pass through the window and get deposited in the detector material. This is only strictly necessary when doing standardless analysis, but the calculations are always done this way for consistency, and to make it easier to compare calibration coefficients between elements. If the theory were perfect, all the calibration coefficients would have the same value. In practice, differences should be relatively small, especially in comparison with coefficients that did not fully compensate for the detector efficiencies. Usually when calibrating elements that all use the same line series (e.g., K), the coefficient variation is small (< 30%), but is often larger when calibrating from mixed lines (e.g., K and L) because it is difficult to make absolute calculations that include the line series information (e.g., fluorescence yields).

It is not necessary to collect pure-element spectra for FP analysis, as no direct ratioing is necessary for the elemental intensities. The calculations are done this way to make it easier to do standardless analysis. Of course, it is possible to use pure-element standards if desired, and the complete FP calibration may be done this way, without any “type” standards being used at all. This is useful if the analyst does not have type standards readily available.

Both direct and secondary fluorescence effects are considered in the FP calculations. Included in the FP database are all the required parameters to calculate or recall absorption coefficients, fluorescence yields, jump factors, Coster-Kronig transitions, line energies, line ratios, etc.

The software consists of a main program that provides the user interface to two libraries (DLL’s) that provide the spectrum processing and FP calculations. It runs on standard PCs and operating systems (Windows 98SE and later), with at least 128MB RAM of memory. Complete XRF analysis is possible, with or without standards, using an internal database of fundamental parameters (FP) such as absorption coefficients, fluorescence yields, transition probabilities, etc. The XRF-FP software is completely compatible and integrated with the Amptek ADMCA display and acquisition software.

Fundamental Parameters Analysis

Elements/Components

Can analyze up to 40 elements, as individual elements and/or compounds. Unanalyzed elements can be specified stoichiometrically bound with an analyzed element (e.g., oxides or carbonates). Elements can be analyzed in one or more compounds within the same analysis. One compound (or element) can be analyzed by difference. Any number of compounds (or elements) can be "fixed." For example, solutions, binders and/or hydrated crystals can be analyzed this way.

General Bulk and Thin-Film Analysis

Any bulk, or single-layer (unsupported) thin-film, sample can be analyzed by either standardless or a calibration-with-standards FP approach. Each analysis may use up to 6 excitation conditions per analysis. Each excitation condition can vary almost any analysis setup, including the kV, acquire time, tube (or secondary) target, detector type, detector or tube filter, source focusing optic, atmosphere (air, vacuum, helium), and spectrum processing (e.g., deconvolution type, background removal, sum & escape peak removal).

Optional software is available to handle multilayer samples up to 6 layers, for simultaneous film thickness and composition analysis using FP with standards (please contact Amptek).

Analysis with or without Standards

Many detectors and windows can be fully modeled. This allows analysis without any standards, with normalization to 100% (or any specified factor). This is only possible when a single excitation condition is used.

When more than one excitation is used, at least one of the elements for each condition must have been calibrated. Calibration factors may be generated using any type of standard (e.g., pure element or analytical "type" standard). A single "type" standard may be used, or the calibration may be done with a different standard for each element, or any combination of standards may be used. If some elements are calibrated and some are not, the latter can use calibration coefficients derived from the former group.

The mass thickness of the sample can either be specified or calculated. If the latter, then the analysis cannot be standardless. Several units are possible for thickness measurement, and the density can be calculated theoretically or specified, in the case of linear thickness calculations. Composition units may be ppm or wt%, with the additional output of atomic and mole percent.

Excitation Sources

Can model different x-ray tube types ("reflection" or transmission), using two analytical models (Pella or Ebel), or by the use of a supplied source spectrum, for complete polychromatic source modeling. Different tube windows and filters may also be included. The tube window can be of any composition (e.g., BeO or glass). Any tube anode element may be specified, as well as the tube (electron) incidence and take-off angles. The kV may range from 3 to 60 kV. Provision is provided for including a transmission efficiency file for use with, for example, polycapillary optics placed between the source and the sample.

Radioisotopes can be used, using a source file describing the relative line ratios. For secondary target excitation, monochromatic excitation is assumed.

Detectors

Various detectors (Si-PIN, SDD, CdTe, Si(Li), and Ge) and windows can be fully modeled. The software has provision for the user to input all the required parameters (e.g., contact material and thickness, dead layer, etc.) associated with these detectors and their windows. For more information on CdTe detector analysis please see this page.

Geometry

The complete system geometry can be specified, including the sample incidence and take-off angles, the source-to-optic and/or source-to-sample distances, the sample-to-detector distance, as well as the environmental factors (see above).

geometry definitions
Figure 2. Geometry angle definitions.

Elements, Lines, and Interelement corrections

Includes full corrections for absorption and both thick and thin-film secondary fluorescence. All possible lines are considered for both excitation and fluorescence. The analysis can be performed for all elements from H through Fm, using K, L or M lines in the energy range from 0.1 keV up to 60 keV.

Spectrum Processing Specifications

Spectrum Calibration

Using two known peaks in the spectrum, the software can calculate the effective gain (eV/channel) and offset (zero shift) for the spectrometer. These factors can then routinely be applied to subsequent spectra prior to other spectrum processing. It is vital that peaks are located at their expected energies, otherwise the spectrum processing cannot function correctly.

Background Removal

This module uses iterative filtering to remove all peaks, leaving behind the spectral background. The background spectrum may be displayed or removed from the original spectrum.

original spectrum before background removal
Figure 3. Original spectrum before background removal.

processed spectrum and removed background
Figure 4. Processed spectrum and background. The blue curve is the removed background.

Escape Peak and Sum Peak Removal

Routines are available to remove, at the user's option, both detector escape and sum (pile-up) peaks.

Click here for information on Escape Peak Corrections for CdTe detectors. This is important when characterizing the output x-ray spectrum of an x-ray tube or performing XRF with a CdTe detector.

80 kev tungsten tube output spectrum
Figure 5. Plot showing a tungsten (W) x-ray tube output spectrum taken with a CdTe detector after processing to remove escape events. The gray trace shows the original spectrum. The green trace illustrates the escape events in the original spectrum. These are subtracted from this original spectrum, then the correct energies are computed (by adding in the energy which escaped). The blue trace shows the corrected escape events, which are then summed with the gray trace. The dark black trace shows the final result of the processing with the events in their correct channels.

Smoothing

A specified number of 1:2:1 Gaussian smoothes can be applied to a spectrum.

Intensity Extraction

Specified element peaks may be integrated over a fixed ROI (Region Of Interest), or the complete spectrum can be fit using synthetic Gaussians for every possible line in the regions of interest. One of six major lines (Ka, Kb, La, Lb, Lg, Ma) is selectable as the main analysis peak for intensity extraction. All relevant lines required for deconvolution are then automatically included by the software for full overlap correction using a least-squares fitting procedure.

All required line energies and resolutions are calculated automatically from the specified analyte line. The Gaussian peak fitting can be done with a linear or non-linear least-squares approach. The latter allows constrained changes in the peak positions, intra-series line ratios, and peak widths, from their nominal starting points.

In addition to calculating elemental intensities, the software automatically calculates the estimated uncertainty and background values, which allows uncertainty and Minimum Detection Limit (MDL) calculations to be performed during the FP analysis.

Software Options

Spectrum Acquisition

There are two methods of spectrum acquisition. The first is to acquire with the Amptek ADMCA acquisition application which controls the MCA8000A, DP4 Digital Pulse Processor, X-123,or PX4 Digital Processor and Power Supply. Once acquired, the XRF-FP can import the file and use the ADMCA display for spectrum processing. The second uses either the MCA8000A control DLL or the DP4/X-123/PX4 control DLL to acquire a spectrum directly into the XRF-FP software which can then be automatically processed. A repeat measurement capability is provided.

xrf-fp auto mode
Figure 6. Auto and Repeat Mode.

Automatic Peak/Element ID

Using the Amptek ADMCA application, peaks can be identified and element tables imported into XRF-FP.


Complete XRF System

complete xrf system
Figure 7. Complete XRF system.

Complete XRF System Includes


X-Ray Fluorescence - A Description

Fp product brochure in PDF format (660 K)
Glossary