xr100cdte x-ray and gamma-ray detector

The use of CdTe diode detectors vs. CZT

Recently developed CdTe Schottky diode detectors1 with 1 mm thickness have shown an advantage over 2 mm thick CZT detectors. The use of Schottky contacts in CdTe detectors reduces the leakage current far below that of CZT. This permits the use of a much higher bias voltage than in a standard detector. Since the charge transport properties of the CdTe detector are much higher than CZT, the net result is a vast improvement in the usable depth (sensitivity). Additionally, because hole tailing and electronic noise are reduced, not only is the sensitivity improved but resolution is much better.

The use of the Rise Time Discriminator (RTD) in the main amplifier has the effect of eliminating pulses that originate from events near the back contact of the detector. Sometimes, more than 50% of the actual thickness of the detectors near the back contact is in effect eliminated through the use of RTD in order to obtain a spectrum with good resolution. The new CdTe diode detectors can produce high resolution spectra without the use of RTD.

cdte 241am spectrum

The Am-241 spectrum above was taken with a 3x3x1 mm thick CdTe-diode detector without the use of RTD. A resolution of 800 eV FWHM was obtained at the 59.54 keV peak.

The spectra below are typical of a 3x3x1 mm CdTe diode detector using rise time discrimination.

All spectra taken with Amptek MCA8000A Multichannel Analyzer.

xr100t-cdte x-ray and gamma ray detector 241-Am spectrum
xr100t-cdte x-ray and gamma ray detector 55-Fe spectrum
xr100t-cdte x-ray and gamma ray detector 109-Cd spectrum
xr100t-cdte x-ray and gamma ray detector 57-Co spectrum
xr100t-cdte x-ray and gamma ray detector 137-Cs spectrum
xr100t-cdte x-ray and gamma ray detector 235-U spectrum
xr100t-cdte x-ray and gamma ray detector CdTe 1 mm thick efficiency curve


Reference

1T. Takahashi and S. Watanabe, "Recent Progress in CdTe and CdZnTe detectors," IEEE Trans. Nucl. Sci. Vol. 48, No. 4, pp. 950-959, Aug. 2001


Amptek has replaced the 2 mm thick CZT with the 1 mm thick CdTe diode detector in order to obtain the best performance. CZT is no longer available.